EasyManuals Logo

Emerson Rosemount 1195 User Manual

Emerson Rosemount 1195
229 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #156 background imageLoading...
Page #156 background image
Rosemount Model 1195/ProPlate/Mass ProPlate
9-2
REMOVE THE
FLOWMETER
FROM SERVICE
NOTE
Once you have determined a that flowmeter is inoperable, remove it
from service.
Be aware of the following:
Isolate and vent the process from the flowmeter before removing
the flowmeter from service.
Remove all electrical leads and conduit.
Do not detach the process flange or the electronics without
consulting the factory.
Symptom Possible Cause Corrective Action
Questionable accuracy or
erroneous flow signal
Improper installation
Is the ProPlate flow arrow pointed in the direction of the flow?
Verify that the cross reservoirs are perfectly level with one another.
Is there sufficient straight run upstream and downstream of the ProPlate?
System leaks Check for leaks in instrument piping. Repair and seal all leaks.
Contamination/plugging Remove the ProPlate and check for contamination.
Closed valve Verify that both ProPlate (PH & PL) or (MH & ML) valves are open. Verify that
vent, equalizer, and line valves are properly positioned per the start up
procedure.
ProPlate calibration
Is the ProPlate calibration too high or low for the flow rate?
ProPlate connections (remote mount only) Verify that the high side of the ProPlate electronics is connected to the high side
of the ProPlate. Check the same for the low side.
Entrapped air (liquid applications) Are there uneven water legs caused by air entrapment in the instrument
connections? If so, bleed air.
Operating conditions Are the operating conditions in compliance with those given at the time the
flowmeter was purchased? Check the flow calc and the fluid parameters for
accuracy. Double-check pipe inside diameter for proper ProPlate sizing.
Spiking flow signal Two-phase flow The ProPlate is a head measurement device and will not accurately measure a
two-phase flow.
Spiking flow signal (Stream
Service)
Improper insulation (Vertical pipes only)
Excessive vibration
Added insulation may be required to ensure that a phase change occurs at the
cross reservoirs.
Check the impulse piping for vibration.
Milliamp reading is zero
Check if power polarity is reversed
Verify voltage across terminals (should be 1055V dc)
Check for bad diode in terminal block
Replace electronics terminal block
ProPlate electronics not in
communication
Check power supply voltage at electronics (10.5V minimum)
Check load resistance (250 ohms minimum)
Check if unit is addressed properly
Replace electronics board
Milliamp reading is low or
high
Check pressure variable reading for saturation
Check if output is in alarm condition
Perform 420 mA output trim
Replace electronics board
No response to changes in
applied flow
Check test equipment
Check impulse piping for blockage
Check for disabled span adjustment
Check electronics security jumper
Verify calibration settings (4 and 20 mA points)
Contact factory for replacement
Low reading/high reading
Check impulse piping for blockage
Check test equipment
Perform full sensor trim (if software revision is 35 or higher)
Contact factory for replacement
Erratic reading for pressure
variable
Check impulse piping for blockage
Check damping
Check for EMF interference
Contact factory for replacement

Table of Contents

Other manuals for Emerson Rosemount 1195

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Rosemount 1195 and is the answer not in the manual?

Emerson Rosemount 1195 Specifications

General IconGeneral
BrandEmerson
ModelRosemount 1195
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals