Pre-use Checkout
104 Instructions for Use Apollo SW 4.5n
Part Number: 9053586, 3rd edition
Leak Man/Spont
Figure 49. Man/Spont Ventilation Leak Test
– Figure 49 shows the components tested in the
Man/Spont leak test.
– This test is performed with sub-atmospheric
pressure.
– Leaks are indicated on the Self Test Results
screen by the red/yellow/green test result
indicator and by posting the leak value if it is over
150 mL/min (Leak (Man/Spont) result in Figure
47).
The Apollo determines the current leakage of the
breathing system and breathing hoses. The system
tolerates leaks of up to 150 mL/min.
Note: For leaks of more than 150 mL/min:
Check the components of the breathing
system and the breathing hoses. Repair any
leaks and repeat the leak test.