System Setup
84 Instructions for Use Apollo SW 4.5n
Part Number: 9053586, 3rd edition
Connecting AC power
Figure 33. Location of Auxiliary Outlets on Back of
Machine
Connecting auxiliary devices
The Apollo has two auxiliary outlets on the back of
the machine (1 in Figure 33). Each outlet is rated 4
amps and is protected by circuit breakers.
1. Connect the external device to an outlet on the
back of the machine
Figure 34. Location of Desflurane Outlet
There is also a dedicated (2 amp) outlet for a
Desflurane vaporizer on the side of the machine,
above the vaporizer mount (1 in Figure 34). This
outlet is protected by safety fuses.
1. Install the Desflurane vaporizer in its mount and
connect it to the outlet on the side of the
machine.
1
2
COM 1
COM 2
IV System
SpO
2
!
Caution!
Consult
Accompanying
Documents
ESD
Seneltivity
!
2
1
CAUTION!
Risk of device failure
In the event of a power failure, any devices
connected to auxiliary outlets will not be powered by
the UPS.
Pay special attention to all power indicators of
connected devices.
CAUTION!
Risk of device failure
If HF surgical devices are connected to the auxiliary
outlets, the leakage current may influence the
electronics of the anesthesia machine causing it to
fail.
Do not connect HF surgical equipment to the
anesthesia machine’s auxiliary outlets.
WARNING!
Risk of device failure
If additional power extension sockets are
connected to the auxiliary outlets, device
internal electronics may be overloaded.
Do not connect additional power adapter
sockets to the auxiliary outlets.