EasyManuals Logo
Home>Dräger>Medical Equipment>Apollo

Dräger Apollo User Manual

Dräger Apollo
310 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #120 background imageLoading...
Page #120 background image
Operation Summary
120 Instructions for Use Apollo SW 4.5n
Part Number: 9053586, 3rd edition
Figure 59. Pushing CLIC Adapter Back into PlaceTo deactivate the CO2 bypass function, swing the
disposable absorber (2 in Figure 59) back into the
system until it engages.
End of operation
To set the Apollo to Standby mode:
1. Press the standby key > <, and press the
rotary knob.
The workstation is now in Standby. The fresh-gas
flow should be turned off.
Figure 60. Apollo Shut Down ScreenTo turn the Apollo off completely (from Standby):
1. Press the main power switch on the front of the
machine.
An acoustic tone sounds, and the shut down
screen shown in Figure 60 is displayed during a
10-second shut down delay.
Note: During the shut down delay, the Apollo can
be restarted immediately by pressing the
main power switch.
2. Make sure the flow control valves are closed.
3. Disconnect the pipeline supply hoses from the
terminal units.
2
WARNING!
Risk of fire
In order to avoid the accumulation of potentially
hazardous oxygen concentrations in the
anesthesia machine or the operating room, all
sources of oxygen must be closed and the
anesthesia machine disconnected from them
when the anesthesia machine is not in use.

Table of Contents

Other manuals for Dräger Apollo

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Apollo and is the answer not in the manual?

Dräger Apollo Specifications

General IconGeneral
BrandDräger
ModelApollo
CategoryMedical Equipment
LanguageEnglish

Related product manuals