Cleaning and Maintenance
234 Instructions for Use Apollo SW 4.5n
Part Number: 9053586, 3rd edition
Removing the passive scavenger system
(optional)
Figure 158. Passive Scavenger System1. Remove the waste-gas hose connected between
the hospital waste-gas disposal system and the
bottom connection on the scavenger
(1 in Figure 158).
2. Remove the scavenger hose connected between
the scavenger and the scavenger connection on
the back of the anesthesia machine
(2 in Figure 158).
3. Remove the scavenger system from the
machine.
4. Prepare the parts for conditioning in a cleaning
and disinfection machine.
Note: For cleaning/disinfection instructions for the
passive scavenger system, refer to its
Instructions for Use.
Removing the endotracheal aspiration
system (optional)
Figure 159. Endotracheal aspiration system bracket and
regulator
1. Remove the suction bottle from the slide mount
(1 in Figure 159).
2. Remove the suction regulator from the bracket
(2 in Figure 159).
3. Remove the endotracheal aspiration system
bracket from the side rail on the side of the
anesthesia machine.
4. If using a disposable suction bottle, dispose of
the bottle with infectious waste.
5. If using a reusable suction bottle, empty the
bottle and dispose of contents with infectious
waste.
WARNING!
Risk of infection
Fluids gathered in the suction bottle may be
infectious and therefore dangerous for hospital
personnel.
Always wear gloves when emptying the suction
bottle.