System Setup
Instructions for Use Apollo SW 4.5n 75
Part Number: 9053586, 3rd edition
Installing the flexible arm for the
breathing bag
Figure 30. Breathing Bag Arm Connection
1. Slide the bag arm assembly onto the breathing
bag port on the side of the breathing system
(1 in Figure 30).
2. Tighten the two thumb screws (2 in Figure 30) to
secure.
3. Attach the 90° fitting to the end of the bag arm
(3 in Figure 30), and attach the breathing bag to
the other end of the fitting.
WARNING!
Risk of patient injury
If incompatible materials are used in the patient
circuit, metabolic products may build up.
Breathing bags used on the Apollo must comply
with current ANSI standards.