Specifications
Instructions for Use Apollo SW 4.5n 285
Part Number: 9053586, 3rd edition
Operating data
Operating voltage 100 to 127 VAC, 50-60 Hz, 12.8 A max.
Power consumption at 110 VAC
Standby (without auxiliary outlets) 1.6 A
Typical (without charging the internal
battery, without auxiliary outlets)
1.8 A
Maximum (with auxiliary outlets) 12.8 A
Power input 200 W typically, max. 1.5 kW with power drawn from auxiliary outlets
Standby 180 W
Typical 200 W
Maximum (with power consumption on
auxiliary outlets)
2.5 kW
Internal battery
Type Lead-gel battery
Sealed, maintenance-free
Backup time with new and fully
charged battery (auxiliary outlets not
supplied)
At least 30 minutes
Maximum 90 minutes (depending on ventilation parameters)
Charging time (to reach full power) At least 10 hours
Charging power Maximum 70 W
Auxiliary outlets 2 outlets with automatic circuit breakers rated at 4 A each,
1 outlet for desflurane vaporizer rated 2 A; combined current for all outlets
is 6 A.
The outlets are isolated against mains to reduce leakage current.
Connection for optional halogen lamp 12 V max. 20 W
Compressed gas supply at pipeline supply inlet
All values apply under ATPD conditions:
Supply pressure for O
2, N2O, and Air 39 psi to 100 psi
(2.7 to 6.9 kPa x 100)
Inlet flows
(each gas, according to ISO 80601-2-13)
12 L/min

Table of Contents

Other manuals for Dräger Apollo

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Apollo and is the answer not in the manual?

Dräger Apollo Specifications

General IconGeneral
BrandDräger
ModelApollo
CategoryMedical Equipment
LanguageEnglish

Related product manuals