Monitoring
158 Instructions for Use Apollo SW 4.5n
Part Number: 9053586, 3rd edition
Note: The CO2
curve can be displayed in gray,
dark gray, red, yellow, blue, or green. This
setting can be configured by an authorized
DrägerService representative.
O
2 concentration
– Curve display
– Numerical display:
–FiO
2 (fractional inspiratory O2 concentration)
–exO
2 (expiratory O2 concentration)
– ΔO
2 (difference between inspiratory and
expiratory O
2 concentration)
– Trend curve for O
2
Anesthetic gas
– Curve display
– Numerical display:
– inAgent (inspiratory gas concentration)
– exAgent (expiratory gas concentration)
– MAC (minimum alveolar concentration)
– Trend curve for anesthetic gases and MAC
Airway pressure
– Curve display (P
AW)
– Numerical display:
– PEAK (peak pressure)
– PLAT (plateau pressure)
– PEEP (positive end-expiratory pressure)
– MEAN (mean pressure) (only on data screen)
– Bar graph
Respiratory flow and volume
– Curve display for flow (insp/exp)
– Numerical display:
– MV (expiratory minute volume)
–V
T (tidal volume)
–VT
INSP (measured inspiratory tidal volume)
– ΔV
T (difference between inspiratory and
expiratory tidal volume)
– Freq. (respiratory rate)
–MV
LEAK (difference between the inspiratory
and expiratory minute volume) (only on data
screen)