Cleaning and Maintenance
Instructions for Use Apollo SW 4.5n 227
Part Number: 9053586, 3rd edition
Overview
This chapter provides complete instructions for the
disassembly and cleaning of the Apollo anesthesia
machine.
Note: Set the Apollo to Standby before
disassembly.
Disassembling components
Observe before disassembly
Disconnecting from the mains
Switch off the device and accessory devices and
remove their mains plugs.
Removing the sample line
Figure 146. Removing the Sample Line1. Disconnect the sample line from the Y-piece and
the fitting on the water trap (see Figure 146).
The sample line is a single-use article which must be
disposed of in accordance with the hospital's hygiene
regulations.
CAUTION!
Risk of gas measurement failure and device failure
Disinfectants can damage the sample gas line and
the diaphragm of the water trap.
Sample gas lines are single-use articles and must
be replaced, not disinfected.