Configuration
Instructions for Use Apollo SW 4.5n 215
Part Number: 9053586, 3rd edition
value configured under Pressure Support will
automatically be adopted when synchronization is
activated during operation. The same also applies
with regard to adopting the value for ΔP
PS, although
this cannot be configured in the Volume Mode,
Volume AF Mode, and Pressure Mode.
Gas supply checks
The settings that can be selected in this menu
determine which gas supplies will be checked during
the self test and normal operation (see 2 in Figure
132).
Note: Only the gas supply defined as being
present in the configuration will be included
in the self test.
Ventilator Default Settings
See 3 in Figure 132.
–P
INSP adjusts with PEEP: Yes or No
When >Yes< is set:
Changes in the set PEEP parameter
automatically changes the parameter value
P
INSP so that the difference between PEEP and
P
INSP remains constant.
When >No< is set:
Parameter value P
INSP remains unaffected by
changes inthe ventilation parameter PEEP.
Factory setting: >Yes<
Connected gas supplies Factory setting
O
2 line: Yes/No Yes
Air line: Yes/No Yes
N
2O line: Yes/No Yes
O
2 cylind.: Yes /No Yes
Air cylind.: Yes/No No
N
2O cylind.: Yes/No No
WARNING!
Risk of device failure
The anesthesia machine does not operate
without at least one oxygen supply.
Either the O
2 pipeline supply or the O2 cylinder
supply must be configured for the O
2 supply.