Ventilation
130 Instructions for Use Apollo SW 4.5n
Part Number: 9053586, 3rd edition
Table 4. Adjustment ranges and factory settings for
Volume Mode
2. Press the button for the parameter to be
changed; its color changes to yellow to indicate
that it is selected.
3. Turn the rotary knob to adjust the parameter to
the desired value, and press the rotary knob to
confirm (4 in Figure 65).
Continue to set the values for the other
parameters.
The parameters that can be set for Volume
Mode are shown in Tab le 4 , along with their
adjustment ranges and factory default values.
Starting the volume-controlled ventilation mode
4. When all the parameters have been preset,
press the rotary knob to start volume ventilation
(4 in Figure 65).
The parameter buttons turn dark green and
display the preset parameter values. The
“Volume” indication in the status field stops
flashing and is displayed continuously. Volume
ventilation begins.
If a volume parameter has to be changed during
ventilation:
– Press the button for the parameter to be
changed, turn the rotary knob to the desired
value, and press the rotary knob to confirm.
Note: Volume Mode can also be started without
presetting values:
– Press the >Vol Mode< key located at the
bottom of the display panel, and press the
rotary knob. Default parameter values are
used.
Note: There is 15-second timeout period for
making ventilation mode changes, with a
5-second audible tone sequence after the
first 10 seconds. If the new setting is not
confirmed within the timeout period, the
current ventilation setting remains in effect.
Ventilation
parameters
Adjustment
range
Factory
settings
1)
1) Site defaults can be set instead.
Pressure limitation
>P
MAX<
[cmH
2O]
10 to 70
min. PEEP
+10
40
Tidal volume
>V
T<
[mL]
20 to 1400
2)
2) Optionally 5 mL to 1400 mL.
600
Frequency
>Freq<
3),4)
[bpm]
3) Depending on the configuration, the inspiratory time (TINSP) can be
automatically changed together with adjustment of the frequency so that the
resultant ratio of inspiration to expiration (I:E) remains constant. Only
applies if trigger = OFF. See the chapter “Configuration”.
3 to 100 12
Inspiratory time
>T
INSP<
4)
[sec.]
4) The resultant ratio of inspiration to expiration (I : E) is also displayed in
parallel.
0.2 to 6.7 1.7
Insp. pause time :
insp. time
>T
IP : TINSP<
[%]
0 to 60 10
>PEEP<
[cmH
2O]
0 to 20
max. PMAX –10
0
Trigger sensitivity
>Trigger<
[L/min]
OFF,
0.3 to 15
3.0
(Press. Supp.)
OFF
(Vol./Press.
Mode)
Pressure Support
>ΔP
PS<
5)
[cmH2O]
5) Optional.
OFF,
3 to 50 5
(Press. Supp.)
max. P
MAX–PEEP OFF
(Vol./Press.
Mode)
Rise time
>T
SLOPE<
[sec.]
0.0 to 2.0 0.0
Age
>Age<
[years]
<1 to 120 40
Ideal body weight
>Weight<
5)
[kg/lbs.]
1 kg to 120 kg,
1 lbs. to 240 lbs.
--