Pre-use Checkout
106 Instructions for Use Apollo SW 4.5n
Part Number: 9053586, 3rd edition
Additional suggestions to isolate compo-
nents of the breathing system for leaks:
Carry out the described measures:
Patient Sample Line Isolation:
1. Remove the sample line for gas measurement
and seal the Luer Lock connection on the
Y-piece.
2. Perform leak test.
Exclude the breathing hoses from the leak test
1. Remove the patient circuit from the breathing
system.
2. Install a leak-free hose between the inspiration
and expiration ports. The breathing bag must be
on the bag arm.
3. Perform leak test.
Isolation of Vaporizers:
1. Remove the vaporizers from the anesthesia
machine.
2. Perform leak test.
Note: The self test does not check for internal
vaporizer leakage; after filling or changing
vaporizers, perform the Standby leak test on
each vaporizer (see page 118).