Specifications
Instructions for Use Apollo SW 4.5n 283
Part Number: 9053586, 3rd edition
Specifications
Ambient conditions
Depending on the type of anesthetic agent delivery
unit used, this data may vary.
General information
Units of measurement for pressure 1 hPa = 1 mbar = 1 cmH
2O
100kPa = 0.1MPa = 1bar = 1kPa x 100
All indicated values and accuracy levels apply at
20 °C (68 °F), 60 % relative humidity, and
1013 hPa (760 mmHg).
The accuracy levels indicated below vary
depending on ambient pressure, temperature,
and relative humidity. If one of the ambient
conditions is changed up to the permissible limit,
the accuracy of the corresponding value can
change by up to 50 %. If more than one of the
ambient conditions are changed, the accuracy
may change by up to 100 %.
All patient-related volumes and flow values have
been standardized to the current ambient
conditions (ATPD).
During operation
Temperature
1)
59 °F to 104 °F
(max. 95 °F for charging the battery)
Air pressure
1)
10.15 to 15.37 psi (700 to 1060 hPa)
Relative humidity
1)
25% to 85% (no condensation)
CO
2 concentration of the ambient air
1)
300 ppm to 800 ppm
Altitude Up to 3000 m (9842 ft)
During storage/transportation
Temperature
1)
–4 °F to 140 °F
Battery: 5 °F to 104 °F
2)
Air pressure
1)
7.25 to 15.37 psi (500 to 1060 hPa)
Relative humidity
1)
25% to 85% (no condensation)
1) Depending on the type of anesthetic agent delivery unit used, this data may vary.
2) The longer-term storage at a temperature outside this range may shorten the life of the batteries.