Cleaning and Maintenance
250 Instructions for Use Apollo SW 4.5n
Part Number: 9053586, 3rd edition
Connecting the endotracheal aspiration
system (Optional)
Figure 171. Endotracheal aspiration system Bracket
and Regulator
The optional endotracheal aspiration system for the
Apollo consists of a suction regulator and a bracket
that attaches to the side of the anesthesia machine.
The bracket is used to hold the regulator and a
suction bottle assembly of the customer’s choice.
1. Attach the endotracheal aspiration system
bracket to the side rail on the left side of the
anesthesia machine.
2. Mount the suction regulator (1 in Figure 171)
onto the bracket.
3. Prepare the suction bottle assembly according
the Instructions for Use provided with the bottle.
4. Install the bottle assembly in the slide mount
(2 in Figure 171) on the bracket.
5. Make all necessary connections between the
suction bottle, suction regulator, and pipeline
vacuum system as specified in the Instructions
for Use provided with the bottle assembly.
Before using on patients on again
1. Re-assemble all equipment, see “System Setup”
on page 57".
2. Check readiness for operation, see “Pre-use
Checkout” on page 89.