Configuration
Instructions for Use Apollo SW 4.5n 209
Part Number: 9053586, 3rd edition
Setting alarm limits
Figure 130. Alarm Limits > Alarm LimitsThe following limits may be configured in the menu
Alarm Limits > Alarm Limits (refer to Figure 130).
Default alarm limits
See 1 in Figure 130.
The high and low alarm limits for patient parameters
can be adjusted within the ranges provided in table.
Two dashes > < indicate that the corresponding
alarm is disabled.
Note: The new default alarm limits are effective
whenever the anesthesia machine is
switched on and after selecting >Restore
default settings< in Standby.
Alarm
Adjustment
range
Factory setting
SpO
2 81 to 99, -- --
[%] 80 to 98 92
Pulse 21 to 250 120
[bpm] 20 to 249 50
etCO
2 1 to 75, -- 50
[mmHg] 0 to 74, -- --
inCO
2
[mmHg]
1 to 10 5
MV 0.1 to 20.0, -- 12
[L/min] 0 to 19.9, -- 3.0
FiO
2 19 to 99, -- --
[Vol.%] 18 to 98 20
P
AW 5 to 99 40
[cmH
2O] 0 to 35 8