EasyManuals Logo
Home>Dräger>Medical Equipment>Apollo

Dräger Apollo User Manual

Dräger Apollo
310 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #241 background imageLoading...
Page #241 background image
Cleaning and Maintenance
Instructions for Use Apollo SW 4.5n 241
Part Number: 9053586, 3rd edition
Assembling the breathing system
Figure 160. Assembling the Breathing System1. Place the bottom section on a flat surface
(1 in Figure 160).
2. Fit the metal valve plate onto the bottom section
(2 in Figure 160).
3. Fit the cover securely on top of the valve plate
(3 in Figure 160).
4. Tighten the five sealing screws a quarter turn
clockwise (4 in Figure 160).
Inserting the flow sensors
1
2
3
4
4
4
4
4
NOTE:
Make sure that all blue rubber seals are
correctly fitted in the bottom section of the
breathing system.
WARNING!
Risk of fire
Residual vapors of easily flammable
disinfectants (e.g., alcohols) and deposits that
were not removed during reprocessing can
ignite when the flow sensor is in use.
Ensure particle-free cleaning and
disinfection.
After disinfection, allow the flow sensor to
air for at least 30 minutes.
Before inserting the flow sensor, check for
visible damage and soiling such as residual
mucus, medication aerosols, and particles.
Replace flow sensors when damaged,
soiled, or not particle-free.

Table of Contents

Other manuals for Dräger Apollo

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Apollo and is the answer not in the manual?

Dräger Apollo Specifications

General IconGeneral
BrandDräger
ModelApollo
CategoryMedical Equipment
LanguageEnglish

Related product manuals