Specifications
Instructions for Use Apollo SW 4.5n 287
Part Number: 9053586, 3rd edition
Electromagnetic compatibility (EMC)
Tested to IEC 60601-1-2
Classification
UMDNS Code
GMDN code
Fresh-gas delivery
as per Directive 93/42 EEC, Annex IX Class II b
Universal Medical Device Nomenclature
System – nomenclature for medical devices
10-134
Global Medical Device Nomenclature –
global nomenclature for medical devices
37710
Settings:
O
2 concentration 21 to 100 Vol.%
Fresh-gas flow, delivery 0 to min. 12 L/min per gas (O
2, N2O, Air)
Fresh-gas flow, electrical measuring 0 to 12 L/min volumetric flow per gas (O
2, N2O, Air)
1)
Standardized to ATPD conditions.
Accuracy ±10% or 0.12 L/min, whichever is greater
Resolution 0.01 L/min (from 0 to 0.2 L/min)
0.02 L/min (from >0.2 to 0.5 L/min)
0.05 L/min (from >0.5 to 1.0 L/min)
0.10 L/min (from >1 - 12 L/min)
1) The maximal permissible fresh-gas flow may be limited depending on the anesthetic agent vaporizer used.