Cleaning and Maintenance
Instructions for Use Apollo SW 4.5n 237
Part Number: 9053586, 3rd edition
Disinfecting/Cleaning/Sterilizing
Choice of disinfectant
Only products suitable for surface disinfection may
be used for disinfection. Testing was performed with
Incidin Extra N and Incidur (wiping), and Gigasept FF
and Korsolex Extra (disinfection by immersion).
The following products are not suitable and should
not be used:
– Compounds containing phenols
– Halogen-releasing compounds
– Strong organic acids
– Oxygen-releasing compounds
In addition to the main agents, disinfectants
frequently also contain additives which may damage
the materials used. The supplier/manufacturer of the
cleaning agent/disinfectant should be contacted if
there is any doubt as to the suitability of a product.
Surfaces
Surfaces of the Apollo, compressed gas hoses, and
cables:
– Wipe off impurities with a damp disposable cloth.
– Disinfect with a wipe disinfectant. (Note the
manufacturer's Instructions for Use.)
WARNING!
Risk of electric shock
Fluids entering the device can damage it
causing malfunctions and endangering the
patient. Wipe the device with only moist, not
dripping wet, objects, e.g. sponges or rags. Do
not place canisters with liquids on or over the
device.
Do not allow any liquids to enter openings in the
device.