Configuration
222 Instructions for Use Apollo SW 4.5n
Part Number: 9053586, 3rd edition
Pulse volume (for SpO2, optional)
0 = off
to
9 = maximum volume
Alarms on/off
This setting is used to enable/disable CO
2 alarms,
optional SpO
2 alarms, and bypass mode (for further
information see page 191 and page 193).
The flow apnea alarm can be configured to be a low-
priority alarm (technical alarm). This setting remains
active until switching to standby and starting a new
case.
This configuration is only possible when the setting
Immediate alarm upon apnea? (patients < 6
years) is set no (see page 213).
Param Settings
Figure 141. Param Settings Configuration ScreenThe following settings can be selected in the Param
Settings configuration screen (refer to Figure 141).
Scaling
–CO
2
–PAW
–Flow
–O
2
The setting is made automatically or by selecting
a pre-set scale.
auto: Automatic adjustment to the next higher or
lower scale after two passes if the scaling frame
is exceeded.
Tidal volume
The setting is made automatically or by selecting
a pre-set scale.
auto: A suitable scale is selected automatically in
accordance with the set age:.
<1 year: 50 mL
1 to 2 years: 150 mL
>2 years to 10 years: 500 mL
>10 years: 1000 mL
Units
CO
2: mmHg, Vol.%, kPa
P
AW: hPa, mbar, cmH2O
Agents: Vol.%, kPa

Table of Contents

Other manuals for Dräger Apollo

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Apollo and is the answer not in the manual?

Dräger Apollo Specifications

General IconGeneral
BrandDräger
ModelApollo
CategoryMedical Equipment
LanguageEnglish

Related product manuals