ECG
SelectingtheleadsforECGanalysis
YoucanchoosewhetherthemonitorperformsanECGanalysisusingsingleleadECG
dataordatafrommultipleECGleads.MultipleECGleadswilltypicallyreducefalse
alarmsandimprovethedetectionsensitivity.However,ifmostleadsarenoisyorlow
amplitude,theSingleleadmodeusingthebestavailableECGleadwillhelp.
Witha3-leadwirecablethesettingisSingleleadandcannotbechanged.Ifthe
measurementmodeischangedfromthe3-leadwiremodeto5-,6-,or10-leadmode,
thesettingchangestoMultilead.
1.SelecttheHRparameterwindow.
2.SelectAdvanced>Page1.
3.SelectanoptionfromtheLeadAnalysislist.Thechoicesare:
●Singlelead:TheEkProalgorithmusesoneoftheleadsI,II,III,orV1forthe
analysis.Forsingle-leadanalysiswhenaVR,aVL,oraVFissetastheECG1
Lead,thefollowingmappingisused:aVR=II,aVL=I,aVF=III.Alsonotethat
theSTvaluesareonlycalculatedforthesinglelead.
●Multilead:TheEkProalgorithmusesthefollowingleads:
■5-leadand6-leadmode:I,II,III,andanyleadassignedtoVa.
■10-leadmode:I,II,III,andV1.
Relearningthepatient’sQRSpattern
DuringECGmonitoring,youmayneedtousetheRelearnQRSfeaturewhena
dramaticchangeinthepatient’sECGpatternhasoccurred.Allowingthemonitorto
learnthenewECGpatterncorrectsfalsearrhythmiaalarmsandheartratevalues,
andrestorestheSTmeasurements.Relearningtakestypically30secondsorless.The
messageRelearning...displaysintheECG1waveformareawhilethemonitorrelearns
theQRSpattern.Duringthistime,arrhythmiadetectionisnotavailable.Ifthemonitor
isnotabletorelearnduetoalowamplitudeQRS,forexample,theArrhythmiapaused
alarmistriggered.Beawarethatifarrhythmiaisturnedoff,theNoisyECGalarmis
triggeredinstead.
1.SelecttheHRparameterwindow.
2.SelectAdvanced>Page1.
3.SelectRelearnQRS.
Automaticrelearningtakesplacewhen:
●Themeasurementmodechangesbetweenthe3-leadmodeandanyotherlead
mode.
●TheECG1Leadselectionischangedinthe3-leadmodeorinsingle-leadanalysis.
●TheValeadselectionischangedinthe5-and6-leadmodes.
●TheLeadAnalysissettingischangedfromMultileadtoSinglelead.
●Thepatientisdischarged/caseisended.
SettingtheprimaryHRsource
TheprimaryheartratecanbecalculatedfromtheECGleads,SpO
2
measurement,
orinvasivepressurewaveform.