Design,environmental,andphysicalspecications
CARESCAPESpO
2
–Nellcor
Operatinghumidityrange5%to95%RH(non-condensing)
Non-operatinghumidityrange5%to95%RH(non-condensing)
Operatingaltituderange-500mto4000m(1075hPato616hPa)
Non-operatingaltituderange-500mto5572m(1075hPato500hPa)
Degreeofenclosureprotection
againstsolidobjectsandwater
IP47
Powerconsumption350mWmaximum
Powerrequirement-input
voltage
5VDC±0.25VDC
Powerrequirement-input
current
70mAmaximum
1mdropDropfrom1minmultipleorientationsontoahardsurface
perthetestinIEC60601-1clause15.3.4.1
ShockWithstandshockinaccordancewithIEC60068-2-27:2008
usingconditionssetinIEC60721-4-7:2001class7M3
Broad-bandrandomvibrationWithstandbroad-bandrandomvibrationinaccordance
withIEC60068-2-64:2008usingconditionssetinIEC
60721-4-7:2001,class7M3
CARESCAPEPressurespecications
Length3.6mor1.8m(11.8ftor5.9ft)
Weightlessthan0.26kg(0.57lb)
Operatingtemperaturerange
0°Cto40°C(32°Fto104°F)
Non-operatingtemperature
range
-30°Cto70°C(-22°Fto158°F)
Operatinghumidityrange5%to95%RH(non-condensing)
Non-operatinghumidityrange5%to95%RH(non-condensing)
Operatingaltituderange-500mto4000m(1075hPato616hPa)
Non-operatingaltituderange-500mto5573m(1075hPato500hPa)
Degreeofenclosureprotection
againstsolidobjectsandwater
IP47
Powerconsumption425mWmaximum
Powerrequirement-input
voltage
5VDC±0.25VDC
Powerrequirement-input
current
85mAmaximum
1mdropDropfrom1minmultipleorientationsontoahardsurface
perthetestinIEC60601-1clause15.3.4.1
2094480-001CARESCAPEONE257