SpO₂
SpO
2
functionaltesters
Youcanverifythefunctionalityofpulseoximetersensorandmonitorwithafunctional
SpO
2
testerbutyoucannotevaluatetheiraccuracywithsuchadevice.Formore
information,refertothestandardISO80601-2-61AnnexFF(Simulators,calibrators
andfunctionaltestersforpulseoximeterequipment).
UsingtheSpO₂measurement
ChangingtheSpO
2
waveformscale
NOTE
GETruSignaltechnologyonly.
1.SelecttheSpO
2
parameterwindow.
2.SelectSetup>Page1.
3.SelectthescalefromtheScalelist:
●AUTO:ThescaleisautomaticallyselectedaccordingtotheIrMod%(infrared
modulationpercentage)receivedfromthemeasurementsource.
●Otherscaleoptionsare2,5,10,20,or50%.
ChangingtheSpO
2
waveformsize
NOTE
MasimoSETtechnologyandNellcor
™
sensorswithOxiMax
™
technologyonly.
1.SelecttheSpO
2
parameterwindow.
2.SelectSetup>Page1.
3.SelectthesizefromtheSizelist.
SelectingtheSpO
2
hemodynamicsweepspeed
NOTE
Thissettingadjuststhewaveformspeedforallofthe
hemodynamicparameters.
1.SelecttheSpO
2
parameterwindow.
2.SelectSetup>Page1.
3.SelectanumericvaluefromtheHemoSweepSpeedlist.
Thesmallerthevalue,theslowerthesweepspeed.
SelectingtheSpO
2
astheprimaryheartratesource
TheprimaryheartratecanbecalculatedfromtheECGleads,SpO
2
measurement,
orinvasivepressurewaveform.
NOTE
Thissettingadjuststheprimaryheartratesourceforallofthe
hemodynamicparameters.
NOTE
HRAlarmsmustbeconguredasSingletoenabletheSpO
2
astheprimaryheartratesource.ThissettingisinCareUnit
Settingsandispasswordprotected.