Parameterspecications
Invasivepressurefrequencyresponse
(waveformlter)
0to12Hzor0to40Hz(-3dB),user-selectable
Invasivepressuresweepspeedoptions
6.25,12.5,25,and50mm/s
Invasivepressurewaveformdisplay
scale
Userandautomatic
Invasivepressurewaveformdisplay
scaleselections
0-10to0-300mmHg,withastepsizeof10mmHg(0.0-2.0to
0.0-40.0kPa,withastepsizeof2.0kPa);orautomaticscalebased
onvalidwaveformvaluesfromlast4secondswithalowerlimit
of-100mmHg(-14kPa)andanupperlimitof350mmHg(48kPa)
andastepsizeof10mmHg(2.0kPa)
Invasivepressuretransducerinterfaces
supported
ArgonMedical,ICUMedical,EdwardsLifesciences,UtahMedical
Invasivepressuretransducerexcitation
voltage
+2.5VDC±1.0%
Invasivepressuretransducersensitivity
5μV/V/mmHg
ReplaceablesensorinterfacecableToolremovablesensorinterfacecable
Invasivepressureanalogoutputsignal
accuracy(gainandoffset)
1V/100mmHg±0.5%±2mmHg,pressure<100mmHg(excludes
transducer)
1V/100mmHg±1.5%±1mmHg,pressure100mmHgorgreater
(excludestransducer)
Invasivepressureanalogoutput
bandwidth
0to40Hz
Invasivepressureanalogoutputdelay
<35ms
Temperaturestandardscompliance
ThesystemwithCARESCAPETemperaturecomplieswithISO80601-2-56:2009-10.
TheCARESCAPETemperaturesensorinterfacecableandcompatibleprobe
accessoriesareTYPECFDEFIBRILLATION-PROOFAPPLIEDPARTsperISO80601-2-56
Clause201.6.
Temperatureperformancespecications
Numberofchannels
2
Temperatureparametersdisplayed
T1,T2
Temperaturerange
0°Cto45°C(32°Fto113°F),deltavalues-45°Cto45°C(-49°Fto
113°F)
Temperaturesystemmeasurement
accuracy
CARESCAPEONEsystemexcludingtemperatureprobes:
●18°Cto45°C(64°Fto113°F):±0.1°C(±0.2°F),ratedoutputrange
●0°Ctolessthan18°C(32°Fto64°F):±0.2°C(±0.4°F),extended
outputrange
Temperatureprobeinstructionsforusespecifytheprobeaccuracy.
Temperaturedisplayresolution
0.1°C(0.1°F)
Temperaturemeasurementunits
°Cor°F