Defaultsettings
SettingDescription
ED
ICUNICUORPACU
RespirationRate
alarmlimits
Selectthe
high/lowalarm
limits.
60/OFF60/OFF
(Infant
prole:
60/15.)
75/2060/OFF
(Perfusion
prole:
OFF/OFF.)
60/OFF
Apnealimit
seconds
Selectthetime
delaybeforean
apneaalarmis
generated.
20s
Temperaturedefaultsettings
T1,T2defaultsettings
SelectMonitorSetup>MainSetup>ParameterSetup>Page2>Temperatures>Setup
toadjustthesesettingsforthecurrentpatient.Thefollowingtableliststhefactorydefault
settings.
SettingDescription
ED
ICUNICUORPACU
T1Measurement
Selectthe
temperaturesite
todisplay.
Enabled
T1LabelSelectthe
temperaturesite
label.
T1
T2Measurement
Selectthe
temperaturesite
todisplay.
Enabled
T2LabelSelectthe
temperaturesite
label.
T2
SelectMonitorSetup>MainSetup>ParameterSetup>Page2>Temperatures>T1,T2
Alarmstoadjustthesesettingsforthecurrentpatient.Thefollowingtableliststhefactory
defaultsettings.
SettingDescription
ED
ICUNICUORPACU
T1alarmlimitsSelectthe
high/lowalarm
limits.
38.5/34
(PediatricandInfant
proles:38/35.5)
37.5/3638.5/34
(PediatricandInfant
proles:38/35.5)
T2alarmlimitsSelectthe
high/lowalarm
limits.
38.5/34
(PediatricandInfant
proles:38/35.5)
37.5/3638.5/34
(PediatricandInfant
proles:38/35.5)
T2-T1alarmlimitSelectthehigh
alarmlimit.
4
T1alarmson/off
Enablealarm
whenlimitsare
violated.
DisabledEnabled
(Perfusion
prole:
Disabled)
Disabled
2094480-001CARESCAPEONE365
CARESCAPE ONE User Manual DRAFT 26 October 2017