Invasivepressures
invasivepressurealarmstab,indicatingthatthealarmpriorityhasbeensetto
EscalatingorLow.
Invasivepressurepracticalities
Invasivepressureparameters
Themeasuredinvasivepressureparametersaresystolic,diastolic,andmean.Pulse
ratecanbemonitoredwithanyarterialsite.CPPisacalculatedvaluethatrequiresa
validICPvalueandavalidarterialsitevalue.
Youcanmonitoruptotwopressures.
Thefollowingtableliststheavailablesitenamesanddisplayedvalues.The“X”after
thesitenamerepresentsthechannelnumber,forexample,Art2:
SiteSitename
CARESCAPEONEdisplayed
values
Generalsitenameforthe
specicinvasivepressure
channels1and2
P1toP2
Mean
Arterial
ArtX
Pulserate,systolic,diastolic,
mean
Centralvenous
CVPX
Mean
Femoralarterial
FemX
Pulserate,systolic,diastolic,
mean
Femoralvenous
FemVX
Mean
Intracranial
ICPX
Mean
Leftatrial
LAPX
Mean
Pulmonaryartery
PAX
Systolic,diastolic,mean
Rightatrial
RAPX
Mean
Rightventricular
RVPX
Mean
Umbilicalarterycatheter
UACX
Pulserate,systolic,diastolic,
mean
Umbilicalvenouscatheter
UVCX
Mean
Intra-aorticballoonpump
NOTE
NotavailableintheNICUsoftwarepackage.
GErecommendsthatthesignalsourceusedtotriggeranIABPshouldbethe
intra-aorticballoonpumpitself.Thisensuresthatthetriggersignaliscompatiblewith
allmodesoftheIABP.AnextrasetofECGelectrodesoranadditionalconnection
fromthearteriallinecanbeconnectedtothemonitortoproducewaveformsonthe
monitor’sdisplayforconsolidatedviewing.
Whenusingthemonitorfortriggering,theIABPtriggersofftheDebSyncportand
usestherstzeroedarterialinvasivepressurechannel.Iftheintra-aorticballoon
pumptriggersoffarterialpressure,theanalogoutputdefaultstothenumericallyrst
2094480-001CARESCAPEONE173