Messages
Message
Location
PossiblecausesSuggestedactions
●SpO2high/SpO2low
●al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
●Checkthepatientstatus.
●Adjustalarmlimitsif
necessary.
●SpO2measurement
removed
●al.areaCARESCAPESpO
2
deviceis
disconnected.
●ConnectaCARESCAPESpO
2
.
device
MessagesrelatedtoNIBPmeasurement
Forinformationregardingalarmprioritiesandescalationtimes,seethesupplemental
informationprovided.
Makesureyouarefamiliarwiththegenericlayoutofthescreen.Thiswillhelpyouidentify
whereonscreenthefollowingmessagesappear.Themessagelocationisindicatedwiththe
followingabbreviations:
●al.area=alarmarea
●param.=parameterwindow
●report=reportview
●wavef.=waveformarea
Message
Location
PossiblecausesSuggestedactions
●Calibrated
●param.
Channelcalibratedsuccessfully.●Waituntilthemessage
disappearsbeforestartinga
measurement.
●Calibrating
●param.
Calibrationofachannelisin
progress.
●Noactionrequired.
●Calibrationerror
●param.
Calibrationhasfailed.●Recalibrate.
●Iftheproblempersists,
contactqualiedservice
personnel.
●Callservice
●param.
Technicalfault.●Contactqualiedservice
personnel.
●CheckNIBP●al.area
Systolicand/ordiastolicresults
missing.
●Checkthepatientstatus.
●CheckNIBPcuffandhoses.
●Repeatthemeasurement.
●Controlmeasurement
●param.
Pressurealarmlimitexceeded.●Allowmeasurementto
complete.
●Checkthepatientstatus.
●Cuffocclusion
●param.
Occlusionduringmeasurement
oroverpressuredcuff.
●Checkthecuff.
●Cuffloose
●param.
Loosecufforcuffhose.●Checkthecuffandcuffhose.
●Cuffoverpressure
●param.
NIBPcuffissqueezedduring
measurementandexceededthe
pressuresafetylimits.
●CheckNIBPcuffandhoses.
●Repeatthemeasurement.
218CARESCAPEONE2094480-001
CARESCAPE ONE User Manual DRAFT 26 October 2017