Safety
Accessorieswarnings
WARNING
Single-useproductsarenotdesignedtobereused.Reuse
maycauseariskofcross-contamination,affectthe
measurementaccuracyand/orsystemperformance,and
causeamalfunctionasaresultoftheproductbeingphysically
damagedduetocleaning,disinfection,re-sterilizationand/or
reuse.
WARNING
Useonlyapprovedaccessories,includingmounts,and
debrillator-proofcablesandinvasivepressuretransducers.
Foralistofapprovedaccessories,seethesupplemental
informationprovided.Othercables,transducersand
accessoriesmaycauseasafetyhazard,damagethe
equipmentorsystem,resultinincreasedemissionsor
decreasedimmunityoftheequipmentorsystem,orinterfere
withthemeasurement.
WARNING
ELECTRICSHOCK.Onlyuseprotectedleadwiresandpatient
cableswiththisdevice.Theuseofunprotectedleadwiresand
patientcablescreatesthepotentialformakinganelectrical
connectiontogroundortoahighvoltagepowersourcewhich
cancauseseriousinjuryordeathtothepatient.
WARNING
Fordetailedinstructionsandinformationregardingsupplies
andaccessories,alwaysrefertotheirowninstructionsforuse.
Cableswarnings
WARNING
CABLES.Routeallcablesawayfrompatient’sthroattoavoid
possiblestrangulation.
WARNING
CABLES.Routeallcablesinsuchawaythattheyarenotunder
thepatienttoavoidtheriskofpossiblepressuresores.
WARNING
PERSONALINJURY.Toavoidpersonalinjurytousersorany
otherpersonsmovinginthevicinityofthecablesortubing,
routeallcablesandtubinginsuchawaythattheydonot
presentatrippinghazard.
WARNING
SAFETYGROUND.Removepowercordfromthemainssource
bygraspingtheplug.Donotpullonthecable.
Debrillationwarnings
WARNING
Donottouchthepatient,table,bed,instruments,modulesor
themonitorduringdebrillation.
WARNING
DEFIBRILLATORPRECAUTIONS.Patientsignalinputslabeled
withtheCFandBFsymbolswithpaddlesareprotectedagainst
damageresultingfromdebrillationvoltages.Toensure
properdebrillatorprotection,useonlytherecommended
cablesandleadwires.Usingothercablesorleadwiresmay
resultindamagetotheequipmentandcompromisepatient
andusersafety.
2094480-001CARESCAPEONE29