Messages
Messagesrelatedtotemperaturemeasurement
Forinformationregardingalarmprioritiesandescalationtimes,seethesupplemental
informationprovided.
Makesureyouarefamiliarwiththegenericlayoutofthescreen.Thiswillhelpyouidentify
whereonscreenthefollowingmessagesappear.Themessagelocationisindicatedwiththe
followingabbreviations:
●al.area=alarmarea
●param.=parameterwindow
●wavef.=waveformarea
Message
Location
PossiblecausesSuggestedactions
●Calibrationfail
●T1Calibrationfail/T2
Calibrationfail
●param.
●al.area
Calibrationofachannelfailed.●Checkconnections.
●Iftheproblempersists,
contactqualiedservice
personnel.
●Checkdevice
●Temperature
malfunction
●param.
●al.area
Hardwareorsoftwarefaultin
CARESCAPETemperature:
●Cablefault
●Dataoutofexpectedrange
●ReconnecttheCARESCAPE
Temperature.
●ReplacetheCARESCAPE
Temperature.
●Iftheproblempersists,
contactqualiedservice
personnel.
●Deviceoverheating
●Movetempdeviceto
acoolerlocation
●param.
●al.area
CARESCAPETemperature
surfacetemperatureistoohigh.
●ChecktheCARESCAPE
Temperatureplacementand
ensurethatitisnotcovered
byanythingthatmightlead
tooverheating.
●ReplacetheCARESCAPE
Temperature.
●IdenticalTempdevice●al.areaTherearetwoCARESCAPE
Temperatures.
●RemoveoneCARESCAPE
Temperature.
●Nosensordetected
●param.
Noprobedetected,ortheprobe
isnotcompatible.Alsowhenthe
temperatureprobecheckfails.
●Checktheprobeand
connections.
●Ensurethatyouareusinga
compatibleprobe.
●T1high/T1low
●T2high/T2low
●al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
●Checkthepatientstatus.
●Adjustalarmlimitsif
necessary.
●T2–T1high●al.areaMeasureddeltavalueisequalto
oroutsidethealarmlimits.
●Checkthepatientstatus.
●Adjustalarmlimitsif
necessary.
●Tempmeasurement
removed
●al.area
CARESCAPETemperature
hasbeenremovedoractive
temperaturechannelbecomes
inactive.
●Checkallconnectionsand
reconnectasrequired.
2094480-001CARESCAPEONE225