Defaultsettings
SettingDescription
ED
ICUNICUORPACU
prole:
180/80)
PR(Fem)alarm
on
Enablealarm
whenlimitsare
violated.
Enabled
(ORsoftwarepackagePerfusionprole:Disabled.
SelectMonitorSetup>MainSetup>ParameterSetup>Page1>ECG>HRAlarms
Arrhythmia>PVC/SVCArrhythmiatoadjustthesesettingsforthecurrentpatient.The
followingtableliststhefactorydefaultsettings.
SettingDescription
ED
ICUNICUORPACU
PVCalarmlimitSelectthehigh
alarmlimit.
10610
PVCalarmonEnablealarm
whenlimitsare
violated.
EnabledDisabledEnabled
SVCalarmlimitSelectthehigh
alarmlimit.
10
SVCalarmonEnablealarm
whenlimitsare
violated.
Disabled
ECGarrhythmiaalarmdefaultsettings
SelectMonitorSetup>MainSetup>ParameterSetup>Page1>Arrhythmia
Prioritiestoadjustthesesettingsforthecurrentpatient.
STdefaultsettings
SelectMonitorSetup>MainSetup>ParameterSetup>Page1>ECG>STtoadjustthese
settingsforthecurrentpatient.Thefollowingtableliststhefactorydefaultsettings.
SettingDescription
ED
ICUNICUORPACU
STAnalysisEnableST
analysis.
On
(Adolescent,Pediatricand
Infantproles:Off)
Off
On
(Adolescent,
Pediatric
andInfant
proles:Off;
Perfusion
prole:Off)
On
(Adolescent,
Pediatric
andInfant
proles:Off)
STalarmsdefaultsettings
SelectMonitorSetup>MainSetup>ParameterSetup>Page1>ECG>ST>STAlarms
toadjustthesesettingsforthecurrentpatient.Thefollowingtableliststhefactorydefault
settings.ThesettingsfortheindividualSTleadsI,II,III,aVR,aVL,aVF,andV1toV6canbe
adjusted.
354
CARESCAPEONE2094480-001
CARESCAPE ONE User Manual DRAFT 26 October 2017