Design,environmental,andphysicalspecications
CARESCAPEECGspecications
Length3.7mor1.9m(12.1ftor6.2ft)
WeightLessthan0.57kg(1.26lb),includeslong10leadwireset
Operatingtemperaturerange
0°Cto35°C(32°Fto95°F)
Non-operatingtemperature
range
-30°Cto70°C(-22°Fto158°F)
Operatinghumidityrange5%to95%RH(non-condensing)
Non-operatinghumidityrange5%to95%RH(non-condensing)
Operatingaltituderange-500mto4000m(1075hPato616hPa)
Non-operatingaltituderange-500mto5573m(1075hPato500hPa)
Degreeofenclosureprotection
againstsolidobjectsandwater
IP47
Powerconsumption625mWmaximum
Powerrequirement-input
voltage
5VDC±0.25VDC
Powerrequirement-input
current
125mAmaximum
1mdropDropfrom1minmultipleorientationsontoahardsurface
perthetestinIEC60601-1clause15.3.4.1
ShockWithstandsshockinaccordancewithIEC60068-27:2008
usingconditionssetinIEC60721-4-7:2001class7M3
Broad-bandrandomvibrationWithstandsbroad-bandrandomvibrationinaccordance
withIEC60068-2-64:2008usingconditionssetinIEC
60721-4-7:2001,class7M3
CARESCAPESpO
2
specications
CARESCAPESpO
2
(TruSignal)
Length3.0mor1.8m(9.8ftor5.9ft)
Weightlessthan0.17kg(0.38lb)
Operatingtemperaturerange
0°Cto40°C(32°Fto104°F)
Non-operatingtemperature
range
-30°Cto70°C(-22°Fto158°F)
Operatinghumidityrange5%to95%RH(non-condensing)
Non-operatinghumidityrange5%to95%RH(non-condensing)
Operatingaltituderange-500mto4000m(1075hPato616hPa)
Non-operatingaltituderange-500mto5573m(1075hPato500hPa)
Degreeofenclosureprotection
againstsolidobjectsandwater
IP47
Powerconsumption375mWmaximum
2094480-001CARESCAPEONE255
CARESCAPE ONE User Manual DRAFT 26 October 2017