9
ECG
ECGcompatibilitylimitations
FordetailedinformationregardingCARESCAPEONE,CARESCAPEParameter,and
accessorycompatibility,seethesupplementalinformationprovided.
ECGsafetyprecautions
ECGwarnings
WARNING
Makesurethattheleadwiresetclipsorsnapsdonottouch
anyelectricallyconductivematerialincludingearth.
WARNING
Whenusinganelectrosurgeryunit,notethatthe
measurementcablesdonotincorporatemeanstoprotect
againstburnsincaseofadefectiveESUreturnelectrode.To
avoidburnsatthemonitormeasurementsites,ensurethe
following:
●PropercontactoftheESUreturnelectrodetothepatient.
●ESUreturnelectrodeneartheoperatingarea.
●Measurementelectrodes,leadwiresandprobesfarfrom
thesurgicalsiteandtheESUreturnelectrode.
WARNING
Thisdeviceisintendedtorecordelectrocardiogramsfrom
surfaceECGelectrodes.Itisnotmeantforpositioning
(oating)temporarypacemakerleadwires,performing
pericardiocentesis,orotherinternalapplications.
WARNING
TheMaximumltermayalterthedisplayedECGmorphology.
DonotmakemeasurementsfromthedisplayedorprintedECG
whenthislterisselected.DisplayedSTvaluesarecalculated
beforeapplyingtheMaximumlteringandmaydifferfrom
valuesmeasuredfromthedisplayedorprintedECG.
WARNING
Whentransitioningfroma10-leadcabletoa6-,5-,or3-lead
cable,selecttheUpdateLeadSetoptiontocleartheLead
offmessagefromthedisplay.
WARNING
Disconnectedelectrodesorlooseelectrodeconnectionscan
leadtomissedcriticalseverityECGalarms.Ifthedevice
reportsLeadsoffafterselectingUpdateLeadSetoption,
alwayschecktheelectrodeconnectionstothepatient.
2094480-001CARESCAPEONE97