Alarmspecications
PRIORITIESANDESCALATIONTIMESMESSAGENOTESLOCATION
infolowmediumhigh
Zeroed
param.
0s
Zeroing
param.
0s
Alarmprioritiesandescalationtimesfortemperature
Thefollowingtableshowsthemessageswiththeirprioritiesandescalationtimes.Inthe
table,themessagelocationisindicatedbythefollowing:alarm=alarmarea;param.=
parameterwindow;andwavef.=waveformarea.Inthefollowingtable,0s=nointended
delay.Formoreinformationonalarmmessages,refertotheuserinformationprovided.
PRIORITIESANDESCALATIONTIMESMESSAGENOTESLOCATION
infolowmediumhigh
Calibrationfail
param.
0s
Checkdevice
param.
0s
Device
overheating
param.
0s
IdenticalTemp
device
alarm
0s
MoveTemp
devicetoacooler
location
alarm
0s
Nosensor
detected
param.
0s
T1calibrationfail
/T2calibration
fail
alarm
0s
T1high/T1low
T2high/T2low
alarm
60s60s60s
T1high/T1low
T2high/T2low
Alarmpriorityset
toEscalating.
alarm
60s120s
T2-T1highalarm
60s60s60s
T2-T1highAlarmpriorityset
toEscalating.
alarm
60s120s
Tempdevice
failure:Call
service
alarm
0s
Temp
measurement
removed
alarm0sifno
patient
case/no
admitted
patient
0sifactive
patient
case/
admitted
patient
2094480-001CARESCAPEONE291
CARESCAPE ONE User Manual DRAFT 26 October 2017

Table of Contents

Other manuals for GE CARESCAPE ONE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE CARESCAPE ONE and is the answer not in the manual?

GE CARESCAPE ONE Specifications

General IconGeneral
BrandGE
ModelCARESCAPE ONE
CategoryMedical Equipment
LanguageEnglish

Related product manuals