Invasivepressures
NOTE
ArtandFeminvasivepressurechannelsonly.Notavailablein
theNICUsoftwarepackage.
1.Selecttheinvasivepressureparameterwindow.
2.SelecttheAdvancedtab.
3.SelectIABPOn.
IABPnowdisplaysintheinvasivepressurechannelparameterwindow.
Settinganarterialinvasivepressure
disconnectionalarm
ThecatheterdisconnectionalarmactivatesifthemeanArt,Fem,orUACpressure
fallsbelow10mmHg(1.33kPa).Thecatheterdisconnectionalarmforeacharterial
pressure(Art,Fem,UAC)isenabledbydefault.IftheAlarmOffAllowedoptionis
enabledinCareUnitSettings,youcanenableordisablethealarmforeacharterial
pressure.CareUnitSettingsarepasswordprotected.Refertothesupplemental
informationprovidedformoreinformation.
1.Selecttheinvasivepressureparameterwindow.
2.SelecttheAdvancedtab.
3.SelectCatheterDisconnect.
Settinginvasivepressurealarmlimits
1.Selecttheinvasivepressureparameterwindow.
2.SelecttheAlarmstab.
3.Selectthetabforthedesiredalarmssetting:
●[Channellabel][Channelnumber]Alarms(e.g.,Art2Alarms):Settingsforthe
selectedinvasivepressurechannel.
●HRAlarms:Settingswhentheheartratealarmsarefromasinglesource.
●PR([Channellabel][Channelnumber])Alarms(e.g.,PR(Art)2Alarms):Settings
whentheheartratealarmsarecalculatedfrommultiplesources.
NOTE
Ifafeatureisnotactive,thealarmlimitsaregreyedout.
YoucansetthemonbyselectingAlarmOn.
4.Setthealarmlimits.
Invasivepressuresalarmpriorities
YoucanselectprioritiesfortheArthigh/low,Femhigh/low,FemVhigh/low,CVP
high/low,PAhigh/low,RAPhigh/low,RVPhigh/low,LAPhigh/low,ICPhigh/low,
CPPhigh/low,UAChigh/low,UVChigh/low,andP1high/lowtoP2high/lowalarms
throughAlarmsSetup>AlarmPriorities>InvasivePressures.Thechoicesare
Escalating,High,Medium,andLow.
EscalatingandLowprioritiesareonlyavailableiftheyhavebeenallowedinCareUnit
Settings.CareUnitSettingsarepasswordprotected.Whenaninvasivepressure
alarmpriorityhasbeensettoEscalatingorLow,awarningtextappearsinthe
172CARESCAPEONE2094480-001