Electromagneticcompatibility
deviceasrecommendedbelow,accordingtothemaximumoutputpowerofthe
communicationsequipment.
Separationdistanceinmeters(m)accordingtofrequencyof
transmitter
150kHzto80MHz80MHzto800MHz800MHzto2.5GHz
Ratedmaximumoutputpowerof
transmitterinwatts(W)d=1.2√Pd=1.2√Pd=2.3√P
0.010.120.120.23
0.10.380.380.73
11.21.22.3
103.83.8
7.3
100121223
Fortransmittersratedatamaximumoutputpowernotlistedabove,therecommended
separationdistance[d]inmeters(m)canbeestimatedusingtheequationapplicable
tothefrequencyofthetransmitter,wherePisthemaximumoutputpowerratingof
thetransmitterinwatts(W)accordingtothetransmittermanufacturer.
At80MHzand800MHz,theseparationdistanceforthehigherfrequencyrange
applies.
Theseguidelinesmaynotapplyinallinstances.Electromagneticpropagationis
affectedbyabsorptionandreectionfromstructures,objectsandpeople.
Compliantcablesandaccessories
WARNING
Useonlyapprovedaccessories,includingmounts,and
debrillator-proofcablesandinvasivepressuretransducers.
Foralistofapprovedaccessories,seethesupplemental
informationprovided.Othercables,transducers,and
accessoriesmaycauseasafetyhazard,damagethe
equipmentorsystem,resultinincreasedemissionsor
decreasedimmunityoftheequipmentorsystem,orinterfere
withthemeasurement.
Minimizingelectromagneticinterference
Electromagneticinterference(EMI)cancauseerraticreadings,cessationofoperation,
orotherincorrectfunctioning.Ifthisoccurs,surveythesiteofusetodeterminethe
sourceofthisdisruption,andtakethelistedactionstoeliminatethesource:
●Turnequipmentinthevicinityoffandontoisolatetheinterferingequipment.
●Reorientorrelocatetheinterferingequipment.
●IncreasetheseparationbetweentheinterferingequipmentandtheCARESCAPE
ONEandCARESCAPEParameters.
TheCARESCAPEONEandCARESCAPEParameterscanradiateradiofrequencyenergy
and,ifnotinstalledandusedinaccordancewiththeseinstructions,maythemselves
causeharmfulinterferencewithothersusceptibledevicesinthevicinity.
2094480-001CARESCAPEONE391
CARESCAPE ONE User Manual DRAFT 26 October 2017