ECG
Settingthearrhythmiacategorytoalarm
DependingonwhathasbeenallowedinCareUnitSettings>Parameters>Allowed
Arrh.Levels,youcanselectdifferentarrhythmiacategoriestoalarm.
1.SelectAlarmsSetup.
2.SelecttheArrhythmiatab.
3.SelectSetup.
4.SelecttheArrhythmiaLevelcategoryyouwanttoalarm:
●Full:Allarrhythmiasalarm.
●Lethal:Onlylethalarrhythmiasalarm.
●Off:Noarrhythmiaalarmsaregenerated.
Settingarrhythmiaalarms
WhilemonitoringECG,youcanadjustthesettingsforarrhythmiaalarmconditions.
1.SelectAlarmsSetup.
2.SelecttheArrhythmiatab.
3.SelectLethalAlarms,VentricularAlarms,orAtrialAlarms.
4.Selectanarrhythmiafromthelist.
5.Selectthearrhythmiaalarm’sAlarmPriority.
6.SelectthecheckboxforCreateSnapshotifyouwishtoactivateanarrhythmia
snapshotcreation.
SnapshotsarenotdisplayedontheCARESCAPEONE,buttheycanbeviewedona
hostmonitorwhentheCARESCAPEONEisconnectedasanacquisitiondevice.
SettingHRforVTach
ThissettingdeterminestheminimumvaluefortheHRtotriggertheVTachalarm.
1.SelectAlarmsSetup.
2.SelecttheArrhythmiatab.
3.SelectLethalAlarms>VTachtab.
4.SelectavaluefromtheHRforVTach/minlist.Theoptionsare90to300beats
perminute,inincrementsof10.
Settingthepauseinterval
Youcansetthetimeintervalbetweenthetwoadjacentbeatsbeforethepausealarm
conditionisannunciated.
1.SelectAlarmsSetup.
2.SelecttheArrhythmiatab.
3.SelectAtrialAlarms>Pausetab.
4.SelectavaluefromthePauseIntervallist.
2094480-001CARESCAPEONE117
CARESCAPE ONE User Manual DRAFT 26 October 2017