B
Compatibility
Compatibility
Devicesandversionsnotspecicallystatedhavenotbeenveriedandvalidatedtocomprise
aconformingsystemwiththeCARESCAPEONE.
Inthefollowingsections,compatibledevicesrefertodevicesthathavebeenveriedtobe
compliantwiththestandard60601–1:2012,3.1edition.
Measurementdevicecompatibility
Identicalmeasurementdevices
WARNING
Donotuseidenticalmeasurementdevicesormeasurement
devicesthatmapameasurementtothesamechannelor
parameterwindow.Ifsuchmeasurementdeviceshave
beenconnected,removethemeasurementdevicethathas
beenmostrecentlyconnected.Youcanalsoremoveboth
measurementdevicesandreconnectthenewmeasurement
deviceafterveseconds.
Thefollowingmeasurementdevicesareconsideredidenticalandshouldnotbeused
simultaneouslyinthesameCARESCAPEONEmonitoringsystem.
MeasurementdeviceSimultaneoususe
CARESCAPEECG
Onlyonepersystem.
CARESCAPEPressure
Onlyonepersystem.
CARESCAPESpO
2
CARESCAPESpO
2
–Masimo
CARESCAPESpO
2
–Nellcor
Onlyonepersystem.
CARESCAPETemperature
Onlyonepersystem.
CARESCAPECO
2
Onlyonepersystem.
2094480-001CARESCAPEONE247