Electromagneticcompatibility
Emissionstest
ComplianceElectromagneticenvironment-guidance
RFemissions
CISPR11
Group1
CARESCAPEONEusesRFenergyonlyforitsinternalfunction.
Therefore,RFemissionsareverylowandarenotlikelytocause
anyinterferenceinnearbyelectronicequipment.
RFemissions
CISPR11
ClassA
Harmonicemissions
IEC61000-3-2
ClassA
Voltage
uctuations/icker
emissions
IEC61000-3-3
Complies
Theequipmentissuitableforuseinallestablishmentsotherthan
domesticandthosedirectlyconnectedtothepubliclow-voltage
powersupplynetworkthatsuppliesbuildingsusedfordomestic
purposes.
Electromagneticimmunity
Thisdeviceisintendedforuseintheelectromagneticenvironmentspeciedbelow.
Itistheresponsibilityofthehospitaltoassurethatthedeviceisusedinsuchan
environment.
U
t
istheACmainsvoltagepriortoapplicationofthetestlevel.
Guidanceandmanufacturer’sdeclarationelectromagneticimmunity
ImmunitytestIEC60601testlevelCompliancelevel
Electromagneticenvironment
guidance
ElectrostaticDischarge
(ESD)
IEC61000-4-2
±6kVcontact
±8kVair
Floorsshouldbewood,concrete
orceramictile.Ifoorsare
coveredwithsyntheticmaterial,
therelativehumidityshouldbe
atleast30%.
ElectricalFast
Transient/Burst
IEC61000-4-4
±2kVforpowersupplylines
±1kVforinput/outputlines
SurgeIEC61000-4-5
±1kVdifferentialmode
±2kVcommonmode
Mainspowerqualityshouldbe
thatofatypicalcommercialor
hospitalenvironment.
Voltagedips,short
interruptionsand
voltagevariationson
powersupplyinput
lines
IEC61000-4-11
<5%U
t
(>95%dipinU
t
)for0.5cycles
40%U
t
(60%dipinU
t
)for5cycles
70%U
t
(30%dipinU
t
)for25cycles
<5%U
t
(>95%dipinU
t
)for5s
Mainspowerqualityshouldbe
thatofatypicalcommercial
orhospitalenvironment.Ifthe
useroftheequipmentrequires
continuedoperationduring
powermainsinterruptions,
itisrecommendedthatthe
equipmentbepoweredfroman
uninterruptiblepowersupplyor
abattery.
PowerFrequency
(50/60Hz)Magnetic
Field
IEC61000-4-8
3A/m
Powerfrequencymagnetic
eldsshouldbeatlevels
characteristicofatypical
locationinatypicalcommercial
orhospitalenvironment.
2094480-001CARESCAPEONE389
CARESCAPE ONE User Manual DRAFT 26 October 2017

Table of Contents

Other manuals for GE CARESCAPE ONE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE CARESCAPE ONE and is the answer not in the manual?

GE CARESCAPE ONE Specifications

General IconGeneral
BrandGE
ModelCARESCAPE ONE
CategoryMedical Equipment
LanguageEnglish

Related product manuals