Parameterspecications
Temperatureprobetypessupported
Series400
Temperaturetestmeasurementcycle
Everyminute
Reusableanddisposabletemperature
probesupport
Sensorinterfacecablesprovidereusableordisposable
temperatureprobeinterfaces.
ReplaceablesensorinterfacecableToolremovablesensorinterfacecable
CO
2
standardscompliance
ThesystemwithCARESCAPECO
2
complieswithISO80601-2-55:2011-12.
TheCARESCAPECO
2
accessoriesthatareintendedtobeconnectedwiththebreathing
systemareTYPEBFDEFIBRILLATION-PROOFAPPLIEDPARTsperISO80601-2-55
Clause201.4.6.
CO
2
performancespecications
CO
2
range0-19.7%
CO
2
accuracy
Aftertwominuteswarm-up:
●Forvaluesbetween0and40mmHg(0and5.3kPa):±2.0
mmHg(±0.29kPa)
●Forvaluesfrom41–70mmHg(5.4–9.3kPa):±5%ofreading
●Forvaluesfrom71–100mmHg(9.4–13.3kPa)±8%ofreading
●Forvaluesfrom101–150mmHg(13.4–20kPa):±10%ofreading
Atrespirationratesabove80rpm,allrangesare±12%ofreading.
ThespecicationsarevalidforgasmixturesofCO
2
,balanceN
2
,
drygasat760mmHg(101.3kPa)withinspeciedoperating
temperaturerange.
Resolution●Numeric:1.0mmHg(0.1kPa)
●Wave:0.1mmHg(0.01kPa)
Stability●Short-termdrift:±0.8mmHg(0.13kPa)oversixhours
●Long-termdrift:Accuracyspecicationismaintainedovera
120-hourperiod
awRR(airwayrespiratoryrate)range
2-150rpm
awRR(airwayrespiratoryrate)accuracy
±1rpm
Warm-uptime
2minuteswithCO
2
sensorattachedforfullaccuracyspecication
Sampleowrate
50±10ml/minute
Totalsystemresponsetime3secondsforon-airwayadapterkits
(Additional30msforsidestreamsamplingcannulas)(Additional2
secondsforextensionlineanddehumidicationtubing)
Totalsystemrisetime200msforon-airwayadapterkits
(Additional30msforsidestreamsamplingcannulas)(Additional80
msforextensionlineanddehumidicationtubing)
CO
2
sweepspeedoptions
0.625,6.25,12.5,25,and50mm/s
2094480-001CARESCAPEONE321