ECG
WARNING
HEARTRATEALARMINTERFERENCE.Poorcablepositioning
orimproperelectrodepreparationmaycauselineisolation
monitortransientstoresembleactualcardiacwaveformsand
thusinhibitheartratealarms.Tominimizethisproblem,follow
properelectrodeplacementandcablepositioningguidelines
providedwiththisdevice.
ECGcautions
CAUTION
Thepatient'sskinmaybecomeirritatedafterprolonged
contactwithelectrodegeloradhesive.
ECGmeasurementlimitations
●ThemonitorwilldisplayaLeadsoffmessageinaninputoverloadcondition,or
upondisconnectionofelectrodeleadwires.
ECGpointstonote
●Thisequipmentissuitableforuseinthepresenceofelectrosurgery,astested
accordingtoIEC60601-2-49clause202.6.2.101Electrosurgeryinterference.
●Forinformationonmaterialsusedinaccessoriesandtheirbiocompatibility,refer
totheinstructionsforuseintheaccessorypackage.
●DebrillatordischargemayaffecttheECGmeasurement.Refertothesupplemental
informationprovidedfordetailsregardingtherecoverytimefromdebrillator
discharge.
●Pre-gelledECGelectrodesarerecommended.Checktheexpirationdate.
●Makesuretheelectrodegelismoist.
●Makesuretheelectrodeshavegoodskincontact.
●Replaceallelectrodesatleastevery24to48hours.
●SelecttheDeactivateECGLeadsOffoptiontoremoveaLeadsoffmessagefrom
thedisplaywhenacableisoff.
ECGmeasurementsetup
ECGequipmenttopatientconnectionwithCARESCAPEONE