Defaultsettings
SelecttheOtherParameterstabtoadjustotherparameteralarmprioritysettingsforthe
currentpatient.Formoreinformation,refertotheuserinformationprovided.Thefollowing
tableliststhefactorydefaultsettings.
Setting
ED
ICUNICUORPACU
SpO2highEscalating
SpO2lowEscalating
SpO2probeoffEscalating
NIBPhigh/lowMedium
RR(Imped.)high/lowEscalating
Apnea(Imped.)Escalating
Resp(Imped.)
measurementpaused
Escalating
RR(CO2)high/lowEscalating
Apnea(CO2)
Escalating
CO2high/lowMedium
Temphigh/lowEscalating
Arrhythmiaalarmsettings
Lethalalarmsettings
SelectAlarmsSetup>Arrhythmia>LethalAlarmstoadjustthelethalarrhythmiaalarm
settingsforthecurrentpatient.
SelectAsyVFibandVTachforallavailablesettings.
Thefollowingtableliststhefactorydefaultsettings.
SettingDescription
ED
ICUNICUORPACU
Asystole>Alarm
Priority
Denethealarm
priority.
High
Asystole>Create
Snapshot
Enableautomatic
snapshotonalarm.
Enabled
VFib/VTach>
AlarmPriority
Denethealarm
priority.
High
VFib/VTach>
CreateSnapshot
Enableautomatic
snapshotonalarm.
Enabled
VTach>Alarm
Priority
Denethealarm
priority.
High
VTach>Create
Snapshot
Enableautomatic
snapshotonalarm.
Enabled
VTachCriteria:
HRforVTach/min
Denethecriteriafor
VTachdetection.
100160100
2094480-001CARESCAPEONE
343