Invasivepressures
2.Checkthatthemonitorrecognizescableconnections(activatesthedisplay)forall
thepressurechannelsusedandthepressurevaluesandappropriatewaveforms
aredisplayed.
Invasivepressuremeasurementonthemonitorscreen
Theinvasivepressurechannellabelsareasfollows.“X”indicatesthechannelnumber,
forexample,Art2.Throughoutthisdocument,whenaninvasivepressurelabelis
followedbyan“X”,the“X”representsthechannelnumber.
WiththeexceptionofPA,morethanonechannelcanhavethesamechannellabel,for
example,Art1andArt2.OnlyonechannelcanbelabeledPA.
Label
Description
ArtX
Arterialpressure
FemX
Femoralarterialpressure
FemVX
Femoralvenouspressure
PAX
Pulmonaryarterialpressure
CVPX
Centralvenouspressure
LAPX
Leftatrialpressure
RAPX
Rightatrialpressure
ICPX
Intracranialpressure
RVPX
Rightventricularpressure
UACX
Umbilicalarterialpressure
UVCX
Umbilicalvenouspressure
P1toP2
Non-specicpressurechannellabels
Usingtheinvasivepressuremeasurement
Invasivepressureanalogoutput
CARESCAPEONEsetstherstavailablearterialchannel(ArtX,FemX,UACX)for
invasivepressureanalogoutput.
●Iftherearenoarterialchannelsavailable,therstzeroedchannelwillbeused.
●Iftherearenozeroedchannelsavailable,aatlinewillbetheoutputoftheIP
Analogoutputchannel.
Aboutzeroingtheinvasivepressuretransducers
●Priortomonitoring,zerotransducersatthepatient’sphlebostaticaxis.Zeroingthe
pressuretransducersisveryimportantforaccuratepressuremeasurements.To
avoidinaccuratemeasurements,youmustzerothepressuretransducers:
■Beforemeasuringinvasivepressures.
■Beforeinitiatingtreatmentchangesreliantuponpressuresdata.
■Whenusinganewtransducerortubing.
■AfterreconnectingthetransducercabletotheCARESCAPEPressure.
2094480-001CARESCAPEONE167
CARESCAPE ONE User Manual DRAFT 26 October 2017