Impedancerespiration
3.SettheRespRate(Imped.)limits.
Apneaalarms’deactivationwiththepauseaudiokey
ApneaalarmscanbedeactivatedwiththepauseaudiokeyiftheAllowalarm
deactivationwiththeAudioPausekeyfor:settingApnea(CO2/Imped.)isenabledin
theCareUnitSettings.Thissettingispasswordprotected.
Formoreinformation,seethesupplementalinformationprovided.
Settingtheapneaalarmdelay
YoucanselecttheapneaalarmdelaybydeningsecondsintheApneaLimitSeconds
setting(3-30seconds).Ifanythingotherthanthedefault(20seconds)isselected,the
selectedsecondsaredisplayedintheparameterwindow.
1.Selecttheimpedancerespirationparameterwindow.
2.SelecttheAlarmstab.
3.SettheApneaLimitSecondswiththearrowselectors.
Enablingtherespirationcardiacartifactalarm
ThecardiacartifactalarmcanbeenabledtodisplaytheCardiacArtifactmessage
whentherespirationrateiswithintoleranceof6.25%oftheECGheartrate.Ittakes
about30breathsbeforeacardiacartifactalarmconditionisdetected.
1.Selecttheimpedancerespirationparameterwindow.
2.SelecttheAlarmstab.
3.SelectCardiacArtifact>AlarmOn.
Respirationalarmpriorities
YoucanselectprioritiesfortheApnea(Imped.),Resp(Imped.)measurementpaused,
andRR(Imped.)high/lowalarmsthroughAlarmsSetup>AlarmPriorities>Other
Parameters.TheavailablechoicesdependonwhathasbeenallowedintheCare
UnitSettings(passwordprotectedsetting).Ifallprioritieshavebeenallowed,you
canselectoneofthefollowing:
●Escalating
●High
●Medium
●Low
●Informational
●Off,forResp(Imped.)measurementpausedonly
Turningofftherespirationmeasurement
1.Selecttheimpedancerespirationparameterwindow.
2.SelectSetup>Page1.
3.SelectRespMeasurement>Off.
130CARESCAPEONE2094480-001
CARESCAPE ONE User Manual DRAFT 26 October 2017