Impedancerespiration
2.SelectSetup>Page2.
3.SelecttheImped.RespSmoothingcheckboxtoenablesmoothing.Deselectit
todisablesmoothing.
Turninginspirationandexpirationmarkersonoroff
Youcanselectwhetherornottheinspirationandexpirationmarkersaredisplayedon
therespirationwaveform.
1.Selecttheimpedancerespirationparameterwindow.
2.SelectSetup>Page2.
3.SelecttheInsp./Exp.Markerscheckboxtoturntheinspirationandexpiration
waveformmarkerson.Deselectittoturnthemoff.
Selectingthewaveformspeed
1.Selecttheimpedancerespirationparameterwindow.
2.SelectSetup>Page2.
3.SelectavaluefromtheSweepSpeedlist.
Thelowerthevalue,theslowerthesweepspeed.
Relearningtherespirationpattern
Ifthepatient’sbreathingpatternchangesaftertheinitiallearningprocesshastaken
place,itmaybenecessarytorelearn.Thereisnorespirationratedisplayedduringthe
relearningprocess.Whenrelearningiscomplete,theRelearningmessagewillclear
andtherespirationratewillbedisplayed.
1.Selecttheimpedancerespirationparameterwindow.
2.SelectSetup>Page1.
3.SelectRelearnRespiration.
Thedetectionthreshold(sensitivity)andthewaveformsizeupdateafterthenew
respirationpatternislearned.
Turningonorofftherespirationratealarm
1.Selecttheimpedancerespirationparameterwindow.
2.SelecttheAlarmstab.
3.SelecttheAlarmOncheckboxtoturnthealarmonforRespRate(Imped.).
Deselectittoturnthealarmoff.
IfyoudeselectAlarmOn,youcannotadjustthealarmlimits.
Thedelaysforrespirationratehighandlowalarmsarecongurablefrom0to30
secondsinincrementsof5seconds.Formoreinformation,seethesupplemental
informationprovided.
Settingtherespirationalarmlimits
1.Selecttheimpedancerespirationparameterwindow.
2.SelecttheAlarmstab.
2094480-001CARESCAPEONE129