Invasivepressures
3.SelectanumericvaluefromtheFilterHzlist.
Thesmallertheltervalue,thegreaterthedegreeoflteringthatoccurs.
Selectingtheprimaryarterialsource
Thearterialpressurechannelselectedastheprimaryarterialsourceisusedasthe
sourcefortheprimaryHR,thesecondaryHR,CPP,andMAP.
1.Selecttheinvasivepressureparameterwindow.
2.SelecttheAdvancedtab.
3.SelectanoptionfromthePrimaryArterialSourcelist.
Theselectionappliestoallinvasivepressurechannelswithanarterialpressure
label(Art,Fem,UAC).
TheAUTOoptiondefaultstotherstzeroedarterialsitethathashadavalidpulse
rateforatleast10seconds.Forexample,ifboththeArt1channelandtheFem2
channelarezeroedandhavevalidpulseratesforatleast10seconds,theAUTO
optionwillbeAUTOArt1becausethatisthelowestvalidchannelnumber.
Iftheselectedprimaryarterialsource’ssitelabelischangedtoanotherarterialsite
label,forexample,ArttoFem,theprimaryarterialsourcelabelwillbeupdatedto
thenewlabel.Iftheprimaryarterialsource’ssitelabelischangedtoanon-arterial
sitelabel,suchasCVP,theprimaryarterialsourceselectionwillchangetoAUTO.
SelectingSmartBP
NOTE
ArtandFeminvasivepressurechannelsonly.Notavailablein
theNICUsoftwarepackage.
SmartBPisanalgorithmthattemporarilydeactivatesthearterialandfemoralalarms
whenitdetectsthezeroingofatransducer,fastushingofthesystem,orblood
draws.ThemessageArtifactdisplaysduringthealarmdeactivation.Whenpulsatile
pressurereturnsand15-20beatsaredetected,numericsaredisplayedandalarms
arereactivated.
1.Selecttheinvasivepressureparameterwindow.
2.SelecttheAdvancedtab.
3.SelectSmartBP.
Compensatingforintra-aorticballoonpump(IABP)
waveformirregularities
WARNING
INCORRECTPULSERATE.BesuretoturnofftheIABPsetting
whenthecardiacassistdeviceisnolongerused.Failuretodo
socouldresultinincorrectpulseratereadings.
CAUTION
PATIENTHAZARD.Ifyouchoosetotriggertheballoonpump
fromthemonitor,contacttheballoonpumpmanufacturer
directlyforinterfacerequirements,astheyvaryamong
manufacturers.Sometriggermodesoncertainballoonpump
devicesmaynotbecompatiblewiththeGEanalogoutput
signal,andusemaycontributetopatientinjuryorsub-optimal
pumpingresults.
2094480-001CARESCAPEONE171
CARESCAPE ONE User Manual DRAFT 26 October 2017

Table of Contents

Other manuals for GE CARESCAPE ONE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE CARESCAPE ONE and is the answer not in the manual?

GE CARESCAPE ONE Specifications

General IconGeneral
BrandGE
ModelCARESCAPE ONE
CategoryMedical Equipment
LanguageEnglish

Related product manuals