Defaultsettings
Currentpatientsettings
Currentpatientsettingsarenon-passwordprotectedsettingsthatareadjustedto
meettheneedsofthecurrentpatient.Thecurrentpatientsettingsarenotsavedto
themonitor’spermanentmemory,butmostcurrentpatientsettingsareconguredto
persistoverawarmstart(CARESCAPEONEisshutdownforlessthan2hours).
Conguringcareunitsettings
Toaccessandcustomizethecareunitsettings:
1.SelectMonitorSetup>Defaults&Service>DefaultSetup.
2.EntertheUsername:clinical.
3.EnterthePassword:ChangeMe.
Thisdefaultpasswordshouldbechanged.
4.SelectCareUnitSettings.
5.ChangethesettingsasneededinAlarms,ParameterAlarms,Admit/Discharge,
Units,andParameters.Allchangesareautomaticallysavedandapplied.
Toreverttothefactorydefaultcareunitsettings,endthecase/dischargethe
patientandselectFactoryDefault>Yes.
Alarmsettings
SelectCareUnitSettings>Alarmstodenethealarmsettingsforthecareunit.The
followingtableliststhefactorydefaultsettings.
SettingDescription
ED
ICUNICUORPACU
ShowAlarmLimitsEnablethe
currentalarm
limitstodisplay
inparameter
window.
Enabled
LatchingAlarmsSelectthelevel
ofalarmpriority
tosounduntil
theeventis
acknowledged.
High
None
AlarmLightEnablethealarm
lighttobeturned
off(0%(Off)
Allowed).
DisabledEnabledDisabled
AudioPauseTime
(min)
Selectthenumber
ofminutesto
pauseaudible
alarms.
2
326CARESCAPEONE2094480-001
CARESCAPE ONE User Manual DRAFT 26 October 2017

Table of Contents

Other manuals for GE CARESCAPE ONE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE CARESCAPE ONE and is the answer not in the manual?

GE CARESCAPE ONE Specifications

General IconGeneral
BrandGE
ModelCARESCAPE ONE
CategoryMedical Equipment
LanguageEnglish

Related product manuals