CO₂
CO
2
troubleshooting
ProblemSolution
CO
2
waveformseemsabnormal.●Checkthepatient:
■Checkforanychangeinpatientcondition.
■Ifthepatienthasacannula,checkforproper
placementofthenasalprongs.
■Ifthepatientisbreathingthroughthemouth,
ensurethatanoral/nasalaccessoryisused.
■Abnormalwaveformscanbeseenifthepatient
istalking.
●Checkthesamplelinesandairwayadapters.
Replaceifnecessary.
Elevatedbaseline.●Checkthepatientstatus.
●ChecktheLoFloSidestreamModule.
●MakesurethepatientisnotbreathingCO
2
.Check
thattheamountoffreshgasissufcientforthe
patient,andcheckthattheCO
2
absorberisin
goodcondition.
●Makesurethesamplecellisproperlyconnected(a
clickishearduponconnection)beforezeroing.
●Iftheelevatedbaselineisseenwhentheaccessory
issamplingroomair,performazero.
Excessivesecretionsinsamplelines.●Checktheorientationoftheairwayadapterport.
●Useaccessorieswithdehumidicationtubing.
●IfthemessageSamplelineblockedappears,
replacethesamplelinesandairwayadapters.
Measurementisnotworking.●ChecktheLoFloSidestreamModuleandadapter
positioning.
CARESCAPECO
2
orLoFloSidestreamModuleisnot
working.
●Contactqualiedservicepersonnel.
Readingsseeminaccurate.●Checkthepatientstatus.
●Checkthatthegasexhaustlineisnotkinkedor
blocked.
●Checkthebreathingcircuitforpossibleleaks.
2094480-001CARESCAPEONE195