ECG
ProblemSolution
Switchtoanotherpacemakerdetectionmode.
Whyisthemonitordouble-countingtheheartrate,
alarmingforalowheartrate,ornotdetecting
pacemakerspikes?
Themonitorisnotdetectingpacemakeractivity.
Causesmayinclude:
Thepacemakerdetectionprogramisturnedoff.
Turniton,repreptheskinandrepositionthe
electrodesifnecessary.RelearnECG.
Thepacemakersignalistooweakforthemonitor
todetect.
TheECGsignalistooweakforthemonitorto
detect.
Themonitorisdetectingatrialpacemakerartifact
ornon-QRSfeaturesasbeats.
Ifthemonitorisalarmingforlowheartrateor
asystole,assesstheQRSamplitude:
VieweachECGleadtoassesstheamplitudeofthe
QRScomplexes.ToensurecorrectHRreadings,
a0.5mVQRSamplitudeisrecommendedfora
normalECGsignal.IftheQRSamplitudedrops
below0.5mVoranabnormalQRSwidthoccurs
(morethan120ms),QRSdetectionmaybe
reduced,leadingtofalseasystolealarms.
Ifnecessary,repreptheskinandrepositionthe
electrodes.
RelearnECG.
Arrhythmiamonitoring
Arrhythmiamonitoringwarnings
WARNING
VFib/VTachshouldnotbeconsideredasubstitutefortheV
Tacharrhythmiaalarm.EffortstolowertheVTachalarmlevel
canresultinmissedventriculartachycardiaalarms.
WARNINGLOSSORDETERIORATIONOFARRHYTHMIADETECTION.
Automatedarrhythmiaanalysisprogramsmayincorrectly
identifythepresenceorabsenceofanarrhythmia.A
physicianmustthereforeinterpretthearrhythmiainformation
inconjunctionwithotherclinicalndings.Pleasetakespecial
noteofthefollowingECGwaveformconditions:
Noisywaveforms.NoisyportionsofECGwaveformsare
typicallyexcludedfromanalysis.Theexclusionsare
necessarytoreducetheoccurrenceofinaccuratebeat
interpretationsand/orrhythmalarms.Iftheexcludednoisy
portionsoftheECGwaveformcontaintruearrhythmia
events,thoseeventsmayremainundetectedbythe
system.
Beatamplitudeandduration.Accuratedetectionand
interpretationofbeatsbecomesincreasinglydifcultas
theamplitudeand/ordurationofthosebeatsapproach
2094480-001CARESCAPEONE115
CARESCAPE ONE User Manual DRAFT 26 October 2017

Table of Contents

Other manuals for GE CARESCAPE ONE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE CARESCAPE ONE and is the answer not in the manual?

GE CARESCAPE ONE Specifications

General IconGeneral
BrandGE
ModelCARESCAPE ONE
CategoryMedical Equipment
LanguageEnglish

Related product manuals