SpO₂
optionsare:2s,4s,8s,10s,12s,14s,or16s.WithGETruSignaltechnology,the
optionsare5sand12s.
1.SelecttheSpO
2
parameterwindow.
2.SelectSetup>Page1.
3.SelectthenumberofsecondsfromtheAveraginglist.
SelectingtheMasimoSpO
2
sensorsensitivitylevel
NOTE
MasimoSpO
2
technologyandMasimosensorsonly.
1.SelecttheSpO
2
parameterwindow.
2.SelectSetup>Page2.
3.SelecttheappropriateSensitivityradiobutton:
UsetheNormalsensitivitysettingfornormalpatientmonitoringpurposes.
UsetheMaximumsensitivitysettingforimprovedpoorperfusionperformance.
UsingtheMaximumsensitivitysettingcanreducetheSpO2probeoffdetection
alarm.Itisrecommendedtousethissettingincareareaswheretheapplication
siteisinspectedfrequently.
Nellcor
sensorswithOxiMax
technologydataaveraging
andupdating
TheNellcorsensorwithOxiMaxtechnologyalgorithmautomaticallyextendsthe
amountofdatarequiredformeasuringSpO
2
andpulseratedependingonthe
measurementconditions.Duringnormalmeasurementconditionsinthenormal
responsemode,theaveragingtimeis6to7seconds.
Duringdifcultmeasurementconditions,whichcanbecausedbylowperfusion,
motion,ambientlight,electrocautery,otherinterference,oracombinationofthese
factors,theOxiMaxalgorithmautomaticallyextendsthedynamicaveragingtime
requiredbeyond7seconds.
Iftheresultingdynamicaveragingtimeexceeds20seconds,alowsignalqualitystatus
isreported,whileSpO
2
andpulseratevaluescontinuetobeupdatedeverysecond.
Asthemeasurementconditionsbecomeevenmoredifcult,theamountofdata
requiredcontinuestoexpand.Ifdynamicaveragingtimereaches40seconds,aCheck
probestatusisreportedandthedevicewillreportdashesduringpulsetimeoutand
pulsesearchcondition.
SelectingtheSpO
2
responsetime
NOTE
Nellcor
sensorswithOxiMax
technologyonly.
Youcanselecttheresponse(averaging)time.Normal(default)istherecommended
setting.
1.SelecttheSpO
2
parameterwindow.
2.SelectSetup>Page2.
3.Selecttheradiobuttonfortheresponsetime:NormalorFast.
146CARESCAPEONE2094480-001
CARESCAPE ONE User Manual DRAFT 26 October 2017

Table of Contents

Other manuals for GE CARESCAPE ONE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE CARESCAPE ONE and is the answer not in the manual?

GE CARESCAPE ONE Specifications

General IconGeneral
BrandGE
ModelCARESCAPE ONE
CategoryMedical Equipment
LanguageEnglish

Related product manuals