EasyManuals Logo

GE CARESCAPE ONE User Manual

GE CARESCAPE ONE
394 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #364 background imageLoading...
Page #364 background image
Defaultsettings
CO
2
defaultsettings
SelectMonitorSetup>MainSetup>ParameterSetup>Page1>CO2>Setuptoadjust
thesesettingsforthecurrentpatient.Thefollowingtableliststhefactorydefaultsettings.
SettingDescription
ED
ICUNICUORPACU
Scale%SelecttheCO
2
scale.
0-6
CO2Sweep
Speed
SelecttheCO
2
sweepspeed.
6.25mm/s
CO2Average
Selectthetime
windowthatis
usedtoselect
thedisplayed
CO
2
value.
10s
FiO2level%Selectthe
levelofFiO
2
usedtomake
compensation
tomeasured
CO
2
values.
21-40
N2Olevel%Selectthe
levelofN
2
O
usedtomake
compensation
tomeasured
CO
2
values.
0-40
SelectMonitorSetup>MainSetup>ParameterSetup>Page1>CO2>Alarms>Page
1orPage2toadjustthesesettingsforthecurrentpatient.Thefollowingtableliststhe
factorydefaultsettings.
SettingDescription
ED
ICUNICUORPACU
EtCO
2
alarm
limits
Selectthe
high/lowalarm
limits(%).
8/3
(NeuroandTrauma
proles:6.0/2.7.)
8/38/3
(Neuro
prole:
6.0/2.7;
Perfusion
prole:8/0)
8/3
(Neuro
prole:
6/2.7)
FiCO
2
alarm
limits
Selectthe
high/lowalarm
limits(%).
3/OFF
FiCO
2
alarms
on/off
Enablealarm
whenlimitsare
violated.
Enabled
(ORsoftwarepackagePerfusionprole:Disabled)
364CARESCAPEONE2094480-001
CARESCAPE ONE User Manual DRAFT 26 October 2017

Table of Contents

Other manuals for GE CARESCAPE ONE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE CARESCAPE ONE and is the answer not in the manual?

GE CARESCAPE ONE Specifications

General IconGeneral
BrandGE
ModelCARESCAPE ONE
CategoryMedical Equipment
LanguageEnglish

Related product manuals