EasyManuals Logo

GE CARESCAPE ONE User Manual

GE CARESCAPE ONE
394 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #303 background imageLoading...
Page #303 background image
Parameterspecications
Subjectscomprisedbothadultmenandwomenandspannedarangeofskin
pigmentations.
SpO₂testmethodsusedtoestablishaccuracy
claimsduringlowperfusion
LowperfusionaccuracyofNellcorOximaxtechnologywith
Oximaxsensors
Specicationappliestomonitoringcableperformance.Readingaccuracyinthe
presenceoflowperfusion(detectedIRpulsemodulationamplitude0.03%-1.5%)was
validatedusingsignalssuppliedbyapatientsimulator.SpO
2
andpulseratevalues
werevariedacrossthemonitoringrangeoverarangeofweaksignalconditionsand
comparedtotheknowntruesaturationandpulserateoftheinputsignals.
LowperfusionaccuracyofMasimoSETtechnology
TheMasimoSETtechnologyhasbeenvalidatedforlowperfusionaccuracyinbench
toptestingagainstBiotekIndex2
TM
simulatorandMasimo'ssimulatorwithsignal
strengthsofgreaterthan0.02%andtransmissionofgreaterthan5%forsaturations
rangingfrom70to100%.Thisvariationequals±1standarddeviation.Plusorminus
onestandarddeviationencompasses68%ofthepopulation.
LowperfusionaccuracyofGETruSignaltechnologywith
TruSignalsensors
GETruSignaltechnologywithTruSignalsensorshavebeenvalidatedforlowperfusion
accuracyinasimulatortestforsaturationrangingfrom70%to100%.Thetestwas
conductedusingaFlukeProSim8simulatorusing0.03%pulseamplitudeandThin
Fingertransmissionsetting.
SpO₂testmethodsusedtoestablishpulserate
accuracy
PulserateaccuracyofNellcorOximaxtechnologywith
Oximaxsensors
Accuracyspecicationswerevalidatedusingmeasurementsofhealthynon-smoking
adultvolunteersduringcontrolledhypoxiastudiesspanningthespeciedsaturation
ranges.Subjectswererecruitedfromthelocalpopulationandcomprisedbothmen
andwomenranginginagefrom18-50yearsold,andspannedarangeofskin
pigmentations.PulseoximeterSpO
2
readingswerecomparedtoSaO
2
valuesofdrawn
bloodsamplesmeasuredbyhemoximetry.Allaccuraciesareexpressedas±1SD.
Becausepulseoximeterequipmentmeasurementsarestatisticallydistributed,about
two-thirdsofthemeasurementscanbeexpectedtofallinthisaccuracy(A
RMS
)range.
AdultspecicationsareshownforOxiMaxMAXAandMAXNsensorswiththepulse
oximeter.
2094480-001CARESCAPEONE303
CARESCAPE ONE User Manual DRAFT 26 October 2017

Table of Contents

Other manuals for GE CARESCAPE ONE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE CARESCAPE ONE and is the answer not in the manual?

GE CARESCAPE ONE Specifications

General IconGeneral
BrandGE
ModelCARESCAPE ONE
CategoryMedical Equipment
LanguageEnglish

Related product manuals